  • DeVilbiss® Suction Unit Instruction Guide

The compressor portion of the DeVilbiss Suction Unit Series (excluding internal rechargeable batteries) is warranted to be free from defective workmanship and materials for a period of two years from date of purchase.


Introduction
Your DeVilbiss Suction Unit is a compact medical suctioning device which has been designed for reliable, portable operation. Because of the small size, light weight and DC operation, the DeVilbiss Suction Unit is ideal for providing suction in the home, in transport with optional DC cord, or if your model has an internal rechargeable battery, the unit can be operated anywhere.
Three container options give the choice between the standard disposable containers or an optional long-term reusable container.
Following the recommended operating and maintenance procedures outlined in this instruction guide will maximize the life of this product.
Set-Up
Internal Filter Cartridge Set-Up Disposable Container w/Internal Filter Cartridge & Splash Guard
1. Fully charge battery for 17 hours (7305P series only).
2. Ensure splash guard is securely attached to inside of lid over filter cartridge.
3. Securely attach lid to container.
4. Insert container into holder.
5. Attach 43/8” tubing from filter cartridge to tubing connector.
6. The patient tubing should be connected to the container lid at the outlet labeled .
7. Please assure that all connections are secure and without leaks before using.
8. Verify that unit is at desired suction level before beginning patient suction.
External Filter Set-Up Disposable Container or Reusable Container with External Bacteria Filter
1. Fully charge battery for 17 hours (7305P series only).
2. Securely attach lid to container.
3. Insert container into holder.
4. Connect either end of the 43/8” reusable tubing or 33/8 ” disposable tubing to the tubing connector then connect the other end to the bacteria filter. Ensure that the clear side of the bacteria filter is toward elbow and container when installing/re-installing. Do not reverse direction of filter.
5. The bacteria filter should then be connected to the 90˚ elbow connection, and the 90˚ connection should be connected to the top of the container lid where it says .
6. The patient tubing should be connected to the container lid at the outlet labeled .
7. Please assure that all connections are secure and without leaks before using.
8. Verify that unit is at desired suction level before beginning patient suction.
How to Operate Your 7305 Series DeVilbiss Suction Unit
Before connecting the unit to the AC adapter or 12V DC cord, make sure that the power switch located on the side of the unit is in the “Off” position. Select power source desired.
LED Explanations: L1- Green– External power supplied to unit from AC power source or DC cord. Illuminated when external power is supplied. L2 – Yellow– Battery is being charged. Light will go out when battery is fully charged. (7305P Series only) L3 – Red– Low battery.
Battery Operation– Verify that your unit has an internal rechargeable battery; factory installed on models 7305P series or provider installed on models 7305D series (installed as an option).
Cleaning Instructions Preparation
1. Shut off unit using power switch and allow vacuum to drop. Disconnect power source from the DC input receptacle on the unit.
2. Disconnect tubing and remove container from holder.
3. Carefully remove lid and empty contents.
Disposable Collection Container with Internal Filter Cartridge The disposable collection container and lid are meant for single-patient use.
1. Remove filter cartridge and 4-3/8” tubing and set aside.
2. Wash container, lid and splash guard in warm water/dishwashing solution. Rinse with clean, warm water.
3. Soak in 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131˚F-149ºF/55˚C-65˚C) solution for 60 minutes. Rinse with clean, warm water and air dry.
Disposable Collection Container with External Bacteria Filter
1. Remove bacteria filter, 3-3/8” tubing, and connection elbow and set aside.
2. Wash jar and lid in a solution of warm water with a mild, liquid detergent (e.g. Dawn or Palmolive) and rinse with clean, warm tap water.
3. Soak in 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131˚F-149ºF/55˚C-65˚C) solution for 60 minutes.
Rinse with clean, warm water and air dry.
Reusable Collection Container with External Bacteria Filter
1. Remove bacteria filter, 4-3/8” tubing, and connection elbow and set aside. Remove o-ring and overflow valve from lid.
2. Wash jar, lid, o-ring, and overflow valve in a solution of warm water with a mild, liquid detergent (e.g. Dawn or Palmolive) and rinse with clean, warm tap water.
3. For single patient use: After washing, disinfect using one of the following methods: a. Soak in 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131ºF-149ºF/55ºC-65ºC) solution for 60 minutes.


DeVilbiss® Suction Unit Instruction Guide

  • Product Code: Learning
  • Availability: In Stock

Related Products

Portable Suction Machine

Portable Suction Machine

A Suction Machine is a device for removing liquids or gases by suction, especially an instrument tha..

Tags: Devillbiss, Suction Machine, Rentals

Quality

THE HIGHEST PRODUCT QUALITY

Warranty On Every Product

Prices

BEST PRICES ASSURED

You Are Purchasing For wholeseller

Support

Dedicated Technical Support Team

+91-9929545206